RIGOL Appendix
2 RSA3000 Performance Verification Guide
Absolute Amplitude Accuracy Test:
Power Meter
Measurement
Value A1
Spectrum Analyzer
Measurement
Value A2
Calculation Result
(A1 - A2)
Limit
Pass/Fail
<0.3 dB
Frequency Response Test (PA Off):
Reference
Value 1
Reference
Value 2
Signal
Generator
Output
Frequency
Power Meter
Measurement
Value A1
Spectrum
Analyzer
Measurement
Value A2
System
Error
[1]
Global
Error
[2]
Frequency
Response
[3]
Limit
Pass/F
ail
100 kHz
100 kHz
to 3.0
GHz: <0.7
dB, <0.5
dB (typical)
3.0 GHz
to 4.5
GHz:<0.9
dB, <0.5
dB (typical)
500 MHz
2 GHz
4 GHz
4.499995
GHz
Frequency Response Test (PA On):
Reference
Value 1
Reference
Value 2
Signal
Generator
Output
Frequency
Power Meter
Measurement
Value A1
Spectrum
Analyzer
Measurement
Value A2
System
Error
[1]
Global
Error
[2]
Frequency
Response
[3]
Limit
Pass/F
ail
100 kHz
100 kHz
to 3.0
GHz: <1.0
dB, <0.5
dB (typical)
3.0 GHz
to 4.5
GHz:<1.2
dB, <0.5
dB (typical)
500 MHz
2 GHz
4 GHz
4.499995
GHz
Note
[1]
: System Error = Power Meter Measurement Value A1 - Reference Value 1
Note
[2]
: Global Error = Spectrum Analyzer Measurement Value A2 - Reference Value 2
Note
[3]
: Frequency Response = |Global Error - System Error|

Other manuals for Rigol RSA3030