RIGOL Appendix
4 RSA3000 Performance Verification Guide
6 dBm
16 dB
14 dB
7 dBm
17 dB
13 dB
8 dBm
18 dB
12 dB
9 dBm
19 dB
11 dB
10 dBm
20 dB
10 dB
11 dBm
21 dB
9 dB
12 dBm
22 dB
8 dB
13 dBm
23 dB
7 dB
14 dBm
24 dB
6 dB
15 dBm
25 dB
5 dB
16 dBm
26 dB
4 dB
17 dBm
27 dB
3 dB
18 dBm
28 dB
2 dB
19 dBm
29 dB
1 dB
20 dBm
30 dB
0 dB
Note
[1]
: System Error = Measurement Value of the Power Meter - (+10 dBm - Attenuation of the Program-controlled
Attenuator) - System Error Reference Value
Note
[2]
: Global Error = Maximum Peak of the Spectrum Analyzer - (+10 dBm - Attenuation of the Program-controlled
Attenuator) - Global Error Reference Value
Note
[3]
: Uncertainty = |Global Error - System Error|
10 MHz Reference Output Accuracy Test:
Frequency Counter
Measurement Result
Output
Accuracy
[1]
Limit
Pass/Fail
<10 MHz × (1 ppm × time since last calibration +
0.5 ppm + 1 ppm)
Note
[1]
: Output Accuracy = |Frequency Counter Measurement Result 10 MHz|

Other manuals for Rigol RSA3030

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Rigol RSA3030 and is the answer not in the manual?

Rigol RSA3030 Specifications

General IconGeneral
BrandRigol
ModelRSA3030
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals