RIGOL Chapter 2 Functions of the Front Panel
2-8 VNA User Guide
Ref Position
Sets the reference position.
When it is set to 5, the reference position is located in the middle of the graticule;
when set to 0, it is at the bottom of the graticule; when set to 10, it is located at the
top of the graticule.
Table 2-8 Ref Position
Note
[1]
: The default values of the reference position for different trace formats are as follows:
⚫ 0: Lin Mag, SWR, Lin Mag(DTF), and Log Mag(DTF)
⚫ 5: Log Mag, Phase, Group Delay, Real, Imaginary, Expand Phase, Positive Phase, and
Return Loss (DTF).
Remarks:
⚫ By default, when "SWR" or "Lin Mag" is selected for the trace format, the
reference position is located at the bottom of the Y axis; for other trace formats,
the reference position is located in the middle of the Y axis.
⚫ When the format is "Smith" or "Polar", the reference position is unavailable. The
Ref Position menu is unavailable.
Scale/Div
Sets the Y scale per division of the selected trace. The scale value is displayed at the
graticule top.
When you select a different trace format, the default Y-axis scale value and unit are
different.
Table 2-9 Scale/Div