How to fix Riken Keiki SD-D58-AC when the power cannot be turned on?
- PparkevelynAug 21, 2025
If the Riken Keiki Security Sensors won't power on, possible causes include incorrect wiring (connect the wiring properly), a removed terminal plate (connect it properly), power supply issues such as abnormalities or momentary blackouts (provide the rated voltage and check the UPS, power supply line filter, and insulation transformer), or cable abnormalities like disconnection or short circuit (check the wiring of the detector head and related devices).