- Press the rotary knob to select this menu point
- Change from “SGL” to “ALL” and press the knob again
- Leave the menu using [ESC]
- Perform a complete alignment:
- Shorten the clamps and press [SHORT]
- Now the short calibration will run for about two minutes.
- Open the clamps (If using e.g. kelvin clamps take care to
have about the same position as it will be for the meas-
urements to come ) and press [OPEN]
- Also this procedure will take up to two minutes
5. Measurement of an unknown coil
Preliminary remarks: Coils containing ferromagnetic core ma-
terials may not give exact results as the core material is speci-
fied for a certain frequency range. This has to be checked be-
fore.
- Switch on the measurement voltage / current monitor:
- Press [SELECT] [2]
- Change to (Vm/Im) using the rotary knob
- Change the value to “ON” and press the knob again
- Leave the menu using [ESC]
- Connect the unknown coil to the test structure
- Change to Z / Phi Measurement by pressing [Z - Φ]
- The phase angle should already be positive
- Now search for the best acquisition frequency
- Press [FREQ]
- Use the rotating knob to change the frequency as long
as Φ is closest to 90°
- Press [L-R]
- Now the nominal inductance is displayed in combination
with the serial resistance
- Check the voltage not to drop down too much (down to
35% of the nominal voltage is fine in this case the nomi-
nal voltage is 1V; see section “LEV” on the left upper dis-
play)
-
PHI and Z over frequency for a standard coil with 132µH / enam. copper wire
5. Measurement of an unknown capacitor
Preliminary remarks: Gold caps can not be measured with the
HM8118 as they are too inert for that kind of measurement.
- Switch on the measurement voltage / current:
- Press [SELECT] [2]
- Change to (Vm/Im) using the rotary knob
- Change the value to “ON” and press the knob again
- Leave the menu using [ESC]
- Connect the unknown capacitor to the test structure
- Change to Z / Phi Measurement by pressing [Z - Φ]
- The phase angle should already be negative
- Now search for the best acquisition frequency
- Press [FREQ]
- Use the rotating knob to change the frequency as long
as Φ is closest to -90°
- Press [C-D]
- Now the nominal capacitance is displayed in combination
with the serial resistance
- Check the voltage not to drop down too much (down to 35%
of the nominal voltage is fine in this case the nominal volt-
age is 1V; see section “LEV” on the left upper display)
6. Additional Information
If the voltage drops down to less than 35% of the nominal
frequency, switch on the constant voltage mode. The cali-
bration done before ist still valid when changing to this
mode.
In the constant voltage mode the impedance of the HM8118
is fixed to 25 , meaning that the output voltage is held con-
stantly to the preset level for devices with an impedance of
more than 25 .
This can improve the results mostly for lower inductance
values. It is important to remark, that the accuracy is re-
duced by the factor of 2 when working in this mode.
It is also possible to acquire the DUT’s behavior using the in-
tended working frequency without tuning. If this is pur posed,
the calibration can also only done for this very frequency us-
ing “SGL” Mode (see “calibration, calibrate all frequencies)
References:
- R&S HM8118 User Manual
- R&S FAQ "How to startup your HM8118 using a specific calibration
- R&S FAQ “HM8118 different measurement result of the same coil
- R&S FAQ “HM8118 tu use external BIAS”
Rohde & Schwarz GmbH & Co. KG
Mühldorfstraße 15 | D - 81671 München
Phone + 49 89 4129 - 0 | Fax + 49 89 4129 13777
www.rohde-schwarz.com

Related product manuals