R&S FSH Module Replacement
1145.5973.82 3.15 E-11
Replacing the RF/IF module
Open the instrument.
Disconnect the flat cable (X8) and remove the tape.
Disconnect the connectors (X7) and (X9).
Remove the RF/IF module by unscrewing (S7).
Remove the binder and BNC connector block:
Loosen the four (S8) screws.
Remove the connector block.
Mount the new RF/IF module.
Mount the binder and BNC connector block with the four (S8) screws.
Connect connectors (X7) and (X9).
Mount the flat cable (X8), including the ferrite.
Fasten the flat calbe with the tape.
Close the instrument.
Perform the manual adjustment.
Refer to Chapter 2 “Adjustment”.
S8
S8
S8
S8
S7
X9
X8
X7
BinderandBNC
ConnectorBlock
Tape
Fig. 3-10 Replacing the RF/IF Module