Remote control commands
R&S
®
RTP
1626User Manual 1337.9952.02 ─ 12
TRIGger<m>:CAN:NDBYtes?
Returns the number of data bytes defined by DLC. DLC and NDB are different in CAN
FD for DLCs > 8.
See also: "Data setup: DLC, NDB, Transfer, Condition, Data min, Data max"
on page 568.
Return values:
<NDBytes> Range: 1 to 64
Increment: 1
*RST: 1
Usage: Query only
TRIGger<m>:CAN:FDATa:DPOPerator <DataPosOperator>
Sets the operator to define an exact position or a data range where the instrument
looks for the specified data pattern.
The setting is available in CAN FD option R&S RTP-K9.
The position can be defined if the data field of the frame is longer than 8 bytes - if
TRIGger<m>:CAN:DLC≥9.
Parameters:
<DataPosOperator> ANY | OFF | EQUal | GETHan | INRange | RANGe
ANY = OFF
The data position is not relevant for the trigger condition.
EQUal | GETHan
Equal, Greater or equal than. These conditions require one data
position to be set with TRIGger<m>:CAN:FDATa:DPOSition.
INRange = RANGe
In range: Set the minimum and maximum value of the range with
TRIGger<m>:CAN:FDATa:DPOSition and TRIGger<m>:
CAN:FDATa:DPTO.
*RST: ANY
TRIGger<m>:CAN:FDATa:DPOSition <DataPosition>
Defines the number of the first data byte at which the data pattern may start.
The setting is available in CAN FD option R&S RTP-K9.
Parameters:
<DataPosition> Range: 1 to 57
Increment: 1
*RST: 1
Protocols