4. In the first dialog of the wizard, select the calibration kit (here: "ZV-Z21") and the test
port connector (here: N 50 Ω (f), corresponding to a male calibration standard), and
click "Next".
The next dialog of the calibration wizard shows that only a single calibration standard
needs to be measured.
5. Click the box "Short (m)..." to initiate the measurement of the connected short stand-
ard.
The analyzer performs a calibration sweep and displays a message box with a pro-
gress bar. After completing the sweep the analyzer generates a short sound and a
green checkmark appears in the checkbox.
Performing a Reflection Measurement