R&S ZVH Spectrum Analyzer Mode (R&S ZVH-K1)
Configuring Spectrum Measurements
Operating Manual 1309.6946.12 - 09 129
► Press the SCALE/AMPT key.
► Press the "RF Att / Amp / Imp" softkey.
► Select either the "Auto Low Distortion" or the "Auto Low Noise" menu item.
The R&S ZVH sets the attenuation according to the table above.
► Select the "Manual: 0 dB" menu item for manual selection of the RF attenuation.
The R&S ZVH opens an input field to set the RF attenuation. You can set the
attenuation from 0 dB to 40 dB in 5 dB steps.
To indicate a manual attenuation, the R&S ZVH puts a red dot at the "Att:"
hardware setting ( ).
4.2.2.6 Using the Preamplifier
To increase the input sensitivity, the R&S ZVH provides an integrated 20 dB
preamplifier after the input mixer.
In the default state of the R&S ZVH, the preamplifer is turned off. If you want to
measure signals with low powers, you can turn it on.
► Press the SCALE/AMPT key.
► Press the "RF Att / Amp / Imp" softkey.
► Select either the "Preamp On" or "Preamp Off" menu item.
The R&S ZVH turns the preamplifier on and off.