R&S ZVH Power Meter (R&S ZVH-K9)
Using the Internal Power Meter (R&S ZVH-K19)
Operating Manual 1309.6946.12 - 09 168
5.3 Using the Internal Power Meter (R&S ZVH-K19)
The R&S ZVH also supports power measurements without using a power sensor. In
that case, you can connect the DUT directly to the R&S ZVH and still perform accurate
channel power measurements.
The screen layout is the same as is described in "Using a Power Sensor" on page 157.
Performing and configuring channel power measurements
The configuration of channel power measurements without a power sensor is similar to
measurements with a power meter.
The following features are available:
Defining the frequency
Zeroing the measurement
Selecting the unit
Defining the reference level
Taking additional loss or gain into account
For more information see "Performing and Configuring Measurements" on page 160.
Defining the channel bandwidth
In addition, you can select the channel bandwidth.
Press the MEAS key.
Press the "Channel BW" softkey.
The R&S ZVH opens an input field.
Enter the required channel bandwidth.
The R&S ZVH performs a measurement on the selected channel. Note that it is not
possible to change the measurement time, resolution bandwidth and frequency span.