R&S ZVH Network Analyzer Mode (R&S ZVH-K42)
Vector Voltmeter (R&S ZVH-K45)
Operating Manual 1309.6946.12 - 09 200
6.4.2.1 Performing Transmission Measurements
Calibrate the measurement for transmission measurements.
Press the CAL key.
Press the "Full 2-Port" softkey.
Perform calibration. For more
information see "Performing
Calibration".
Reconnect the DUT.
The R&S ZVH shows the
characteristics of the DUT
(magnitude and phase) as
numerical values.
6.4.2.2 Performing Reflection Measurements
Calibrate the measurement for reflection measurements.
Press the CAL key.
Press the "Full 1-Port" softkey (for calibration on port 1).
Perform calibration. For more information see "Performing Calibration".
Reconnect the DUT.
The R&S ZVH shows the characteristics of the DUT (magnitude and phase) as
numerical values.
Note that for calibration to remain valid, frequency parameters, bandwidth and
attenuation have to remain the same.