R&S ZVH Cable and Antenna Test Mode
Performing Cable and Antenna Measurements
Operating Manual 1309.6946.12 - 09 47
Press the CAL key.
Press the "Full 2-Port" softkey.
The R&S ZVH asks you to connect
an "Open" to port 1.
Firmly connect the "Open" of the calibration standard to port 1.
You can abort the calibration any time by pressing the "Cancel" softkey.
Press the "Continue" softkey to
start calibration.
The R&S ZVH calibrates the
"Open".
Disconnect the "Open".
Next, the R&S ZVH asks you to
connect a "Short" to port 1.
Firmly connect the "Short" of the calibration standard to port 1.
Press the "Continue" softkey to start calibration.
The R&S ZVH calibrates the "Short".
Disconnect the "Short".
Next, the R&S ZVH asks you to
connect a "Load" (50 Ω termination) to
port 1.
Firmly connect the "Load" of the calibration standard to port 1.
Press the "Continue" softkey to start calibration.
The R&S ZVH calibrates the "Load".
Disconnect the "Load".
Next, the R&S ZVH asks you to set up
a "Through" connection from port 1 to
port 2.
Firmly connect the "Through" connection to port 1 and port 2.
Press the "Continue" softkey to start calibration.
The R&S ZVH calibrates the "Through" connection.
After finishing the calibration routine, the R&S ZVH shows that calibration is
finished for a short time ( ). The status line now says (Cal) to indicate
successful calibration.