R&S ZVH Cable and Antenna Test Mode
Analyzing Measurement Results
Operating Manual 1309.6946.12 - 09 61
2.3.1.3 Working with Memory Traces
The R&S ZVH can transfer a trace to the trace memory and also display the current
trace and the trace in the trace memory for comparison. The saved trace is always
displayed in white to distinguish it from the current trace.
Press the TRACE key.
Press the "Trace Memory" softkey.
The R&S ZVH transfers the trace to the trace memory.
Press the "Show Memory" softkey.
The R&S ZVH displays the saved trace in white.
You can remove the memory trace by pressing the "Show Memory" softkey again.
The memory traces are bit-mapped into the picture memory. Therefore, when the
memory trace is recalled, it will not be adapted to any modifications of the reference
level or span that may have been made in the meantime.
When you load a data set, the R&S ZVH stores the corresponding trace in the trace
memory. You can view that trace with the "Show Memory" softkey.
2.3.1.4 Using Trace Mathematics
For more information see "Using Trace Mathematics" on page 144.