Rotek 811AE - 4.2.6 Deviation Test

Default Icon
105 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
26
Configure the equipment for power measurements as in Figure 4-4. Set the Series 800 to
120V, 600VA, 400Hz. Use a DVM to read the MSB-001A DC voltage output. Use the percent
Deviation program of the DVM. At each PF setting "Enter" the DVM reading with the 800 Deviation
control "Off". Then assert the 800 Deviation range and setting. The DVM reading must be between
the limits given.