2 Technical specification
Line voltage: 83 V – 264 V ac, 50 / 60 Hz
Power consumption:
50 VA
Safety class:
1
ambient temperature
:
-
+40 °C
Storage temperature: -10 - +50 °C
Measuring ranges
HF current : 0 - 5000 mA
HF
-
Peak Current (V2):
0
-
5000 mA
Discrimination: 0,1 mA
or HF output power:
0
-
500 Watt
Crest Faktor (V2)
1
-
10 (at > 1000 mA)
HF leakage current:
0
-
250 mA
Discrimination: 0,1 mA
Neutral electrode test: 0 - 1000 Ohm
Range:
0,3
-
10 MHz
Measuring principle: Thermal electric converter
Load resistors: 10 Ohm,
25 Ohm - 6375 Ohm
in steps of 25 Ohm
Swing in time: < 3 sec.
Rated voltage
:
m
ax.
1000 V
Output power: 500 watt 1 min on, 5 min off
Permanent max. 200 watt at 25°C
environmental temperature (50
800 Ohm)