Tissue Processing
Tissue-Tek® Xpress® x120
Quick Reference Guide
Tissue-Tek® Xpress® x120
Sakura Finetek Europe B.V. | Flemingweg 10a, 2408 AV Alphen a/d Rijn | NL
Phone: +31 (0) 88 592 00 00 | Fax: +31 (0) 88 592 00 01 | sakura@sakura.eu | www.sakura.eu
Start Up
The x120 can be programmed to automatically
initiate a four-hour system warm up so it is ready
for operation when the workday begins.
If the PPS On-Board function software is used make
sure the User Guide x120 On-Board PPS Timer
function is used as leading for System Operation.
Loading station filled with Pre-
Processing Solution or
Molecular Fixative.
Basket, holds up to 40
cassettes
Run preparation for processing
Pre-Start Checks
1. Loading Station is filled with Pre-Processing
Solution and replaced every 4 hours, or filled
with Molecular Fixative and refreshed daily
depending on the software version.
2. No baskets are loaded in the instrument.
3. Transfer arm is free for moving.
4. All retort lids are closed.
5. The basket holds up to 40 cassettes but should
not exceed the maximum limit rim.
6. Do not use cassettes that contain any metal
which can interfere with microwave; it can
cause small burning artifacts in the tissue, and
can as well damage the microwave oven.
Access permitted Icon and loading indicator is illuminated

Other manuals for Sakura Tissue-Tek Xpress x120

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Sakura Tissue-Tek Xpress x120 and is the answer not in the manual?

Sakura Tissue-Tek Xpress x120 Specifications

General IconGeneral
BrandSakura
ModelTissue-Tek Xpress x120
CategoryComputer Hardware
LanguageEnglish