Tissue-Tek® Xpress® x120
Tissue Processing
Quick Reference Guide
Tissue-Tek® Xpress® x120
page 10 | 10 QRG | x120 | 2010 | V-10
Do’s and Don’ts
DO NOT use blue sponges for tissue processing.
DO use embedding bags or biopsy cassettes (no mesh cassettes)
DO NOT use metal lids on cassettes for processing
DO run RNA tissue after the processor is cleaned
DO NOT process tissue thicker than 2 mm on the Standard program.
DO process tissue that is trimmed to 3 mm thick using the Extended program.
DO NOT allow the faced tissue blocks to soak in water.
DO face the block and allow the block to cool on ice.
DO use plastic lids and remove metal staples and sutures from tissue prior to placement in the
cassette.
DO NOT top up reagents or rotate reagents. Reagents to be replaced as a set.
DO change the reagents and solution after 1500 cassettes / 35 runs (Reagent Control) or every
40 hours of operation, whichever occurs first. Invert reagent bottles 5 times before use as stated
on the label.
DO follow maintenance and record (initial) on the checklist provided

Other manuals for Sakura Tissue-Tek Xpress x120

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Sakura Tissue-Tek Xpress x120 and is the answer not in the manual?

Sakura Tissue-Tek Xpress x120 Specifications

General IconGeneral
BrandSakura
ModelTissue-Tek Xpress x120
CategoryComputer Hardware
LanguageEnglish