Caution: System must be unlocked before self-test, i.e. set LCK parameter in internal parameter list to
zero. Otherwise there will be no response when pressing ‘cal’ key and Shift key. This function is to prevent
on-the-spot operators from Misoperation. Parameter for measurement range RN can either be set before
self-calibration or after that. Zero pressure must be ensured for induction head of transducer when
self-calibrating.
Attentions:
1 Transducer head must be free from collision and properly screw its protection cap when being
stored;
2 The instrument should be kept in humidity-free environment;
3 The instrument should work with rated voltage, otherwise it may result in unusual failures;
4 The instrument has functioned of temperature zero tracking. Generally, it can zero automatically.
But if temperature drift is so wide that it exceeds the automatic tracking range, press AC key and
point press Shift key simultaneously to calibrate zero pressure. If zeroing by accident, press AC
key for more than 5 seconds and zero calibrated in normal temperature is recovered.
Maintenance:
1 Numerical display ’88.88 or ‘--‘indicates broken transducer leads or improper contact. Connect
properly.
2 No numbers displayed: check if 220 V AC power is connected. If not, check if there is output
from switch power supply.
3 Unstable numbers displayed: generally, it means that ground wire is improperly connected with
ground. Connect shielded wire properly with ground.
4 If major failures occur due to misoperation in connecting wires by clients, please contact us for
replacement of parts such as power board, CPU board, or display panel.
5 If repair work is beyond customers’ capability, please directly send to us. We ensure one-year
responsibility of repair and life-long maintenance.
Ex-factory supplementary components
Serial
number
name unit Quantity remark
1
PS1016 Intelligent Temperature
Pressure Gauge
set 1
2 Transducer piece 1
In term of the
customers request
3 Cable wire meter 3 One end with socket
4 Installation frame piece 2
5 Operation Manual sheet 1
6 Pass certificate sheet 1