EasyManua.ls Logo

Sanyo MBR-1404G - Control Panel Overview

Sanyo MBR-1404G
26 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
12
REFRIGERATOR COMPONENTS
Control panel
1. Alarm buzzer stop key (BUZZER): Press this key to temporarily silence the buzzer in the event that
the alarm operates and buzzer sounds.
2. Temperature display (UPPER/LOWER): By pressing this key, the displayed temperature is
changed; upper monitor bottle temperature, lower monitor bottle temperature, and average temperature
of upper and lower monitor bottles are displayed in sequence.
3. Alarm test key (ALARM TEST): To check the operation of alarm buzzer, lamp, and remote alarm
terminal.
4. Calibration key (CAL): To calibrate the temperature. Normally this key is not used.
5. Digit shift key ( ): Pressing this key in the CAL mode will cause the changeable digits to move.
6. Numerical value shift key ( ): Pressing this key in the CAL mode will cause the numerical value
to change.
7. Door check lamp (DOOR): This lamp is lit when the door is ajar or open.
8. Alarm lamp (ALARM): This lamp is flashed during alarm condition.
9, 10. Temperature display monitor lamp (UPPER, LOWER): Lamp 9 is lit when upper monitor bottle
temperature is displayed. Lamp 10 is lit when lower monitor bottle temperature is displayed. Both 9
and 10 are lit when the average temperature is displayed.
11. Digital temperature display (TEMPERATURE
o
C): The upper or lower monitor bottle temperature
or average temperature is displayed.
7
1 2 3 4 5 6
8 9 10 11

Related product manuals