6 Installation and Starting Up
Electronic positioning valve – 1. During assembly of the valve body, the electronic positioning valve must be
Valve body installation in the position “100%”.
2. Mount the Valve body to the connector (see Figure 3).
3. Fasten the Valve with Tri-clamp (see Figure 4).
4. The connection in between the actuator and the valve body is established
by moving the actuator into the position ”0%”.
Figure 3 Figure 4
Flow sensor 1. The sensor mounted on the flow cell as shown in Figure 5.
Figure 5
Installation of the Pump element 1. Lay the double y-tube element into the open pump head and place the branches
over the fixing screws.
2. Close the lid. It might be helpful to use a wrench (SW10).
Figure 6