Calibration/Adjustment Sequence
You can configure the operating menu
so that:
adjustment automatically follows calibra-
tion in a single operation (1 10 1), or
the operator chooses whether to end
the calibration/adjustment routine or
have adjustment performed (1 10 2)
If there is no deviation, or if the
difference is within the effective require-
ments for accuracy of measurements,
it is not necessary to adjust the balance/
scale. In this case, you can end the
calibration routine following calibration.
Two keys are active at that point:
q = start adjustment
c = end the calibration/
adjustment routine
Factory Settings
Calibration adjustment mode for
models without built-in motorized
calibration weight:
External calibration (1 9 1)
Calibration adjustment mode for models
with built-in motorized calibration weight:
Internal calibration (1 9 3)
Calibration/adjustment sequence:
Adjustment automatically follows calibration
in a single operation (1 10 1)
Weight unit for calibration: grams (1 11 1)
ISO/GLP-compliant printout: off (8 10 1)