Example: Determining residual weight in percent
Parameter settings: Applic.: Percent (menu code 2. 4.)
Reference percentage: Ref 100%
Step Key (or instruction) Display/Data output
1. Tare the balance/scale
w 0.0 g
2. Information:
Enter reference percentage
b ref 100
%
(Changing the reference:
see the previous page)
3. Place sample equal to 100%
on the balance/scale
(in this example: 111.6 g)
4. Initialize the balance/scale
u + 100.0
%
*
The current reference weight pRef 100 %
remains stored until a new reference Wxx% 111.6 g
is set or the power supply
is interrupted
5. Remove sample
(e.g., for drying)
6. Place unknown weight on +
94.9
%
*
balance/scale
(in this example: 322.5 g)
7. If desired, print percentage
r Prc + 94.9 %
8. Toggle display between + 105.9 g
*
weight and percentage Repeatedly: b + 94.9
%
*
9. Clear display of residual weight
and reference percentage
c + 105.9 g
10. If desired, print net residual weight
r N + 105.9 g