32 EVO1Y | LAB1Y Operating Instructions
Faults
8 Faults
Problem Cause Remedy
No segments appear on
the weight display
No AC power is available Check power supply
The weight readout
shows “LOW”
No weighing pan on
the scale
Position the weighing
pan
The weight readout
shows “HIGH”
Weighing capacity
exceeded
Unload the scale
The weight readout
changes constantly
Unstable ambient
conditions
Excessive vibration or
draft
Weighing pan is being
affected at some point
by outside influences
Set up scale in
another area
Adjust the scale
settings (see Chap-
ter “3.4 Menu List,”
page 12)
The weight readout is
obviously incorrect
The sample is not stable
Scale not tared before
weighing
Tare before weighing
No weight value is shown
and the padlock symbol
H is displayed
PC connection to the
scale has been inter-
rupted activating the
“LOCK” function
Manual “LOCK” function
is active
Access the menu to
make the necessary
adjustment and switch
off the “LOCK”
function
Check the connection
Switch off the manual
“LOCK” function