  Installation 
4.4.2  Connecting the Scale
Connecting the Scale
 t Insert the USB cable plug into the USB socket on the back of the display.
 
 t Loosen the screw (1) on the lock.
 t Swivel the lock over the USB cable.
 t Re-tighten the lock screw.
Laying the USB Cable
 t Lay the USB cable through the cable holders on the back of the scale.
Connecting the Grounding Cable
This explosion-protected system should be set up according to commonly accepted 
technical standards. The applicable national electrical code and safety regulations for 
your particular country must be observed.
Before commissioning the scale, a check must be carried out by or under the 
supervision of a qualified electrician to ensure that the system is in good working 
order.
Check whether or not the competent authorities (e.g., industrial supervisory board) 
need to be informed. It is also necessary to carry out inspections of the system 
during operation.
Inspection intervals should be such that any significant defects that may occur can 
be identified in good time. Inspections should be carried out at least once every 
three years. The applicable requirements and guidelines should also be observed 
during operation.
Establish a low resistance connection from the scale to a customer-supplied 
equipotential bonding conductor connection via the equipotential bonding 
conductor connection on the device using a suitable grounding cable with a 
gauge of at least 4 mm
2
 (not included).