  Technical Specifications
11  Technical Specifications
11.1  General Data
Specification Unit Value
ID code (explosion protection)  II 3G Ex ic nA IIB T4 Gc as per EC Type Examination Certificate no. 
FM15ATEX0008X
Power supply Only via USB interface or Sartorius AC adapter YEPS01-USB
Input supply voltage V
DC
+4.5 to 5.0
Power consumption W 2.0 (typically)
Other data IP40 in accordance with EN 60529 / IEC 60529
Ambient Conditions
The specifications apply under the following ambient conditions:
Environment For indoor use only
Ambient temperature*  +10°C to +30°C
Operational capability °C Guaranteed between +5 and +40
Storage and shipping °C -10 to +60
Relative humidity** % 15% to 80% for temperatures up to 30°C non-condensing,  
decreasing linearly to 50% relative humidity at 40°C
Electromagnetic Compatibility In accordance with EN 61326-1/IEC 61326-1 Electrical equipment 
for measurement, control and laboratory use – EMC require-
ments – Part 1: General requirements
Interference resistance Basic requirements
Transient emissions Class B 
Suitable for use in residential areas and areas that are connected 
to a low voltage network that also supplies residential buildings.
Verified scales in accordance with EU requirements comply 
with the requirements of Council Directive 2014/31/EC with 
EN 45501:2015 and OIML R76:2006.
*   For verified scales in accordance with EU requirements,  
refer to the information on the scale.
**  For verified scales in accordance with EU requirements,  
the legal regulations apply.
Available Application Programs Recalculation, factor calculation, formula
 
Power supply YEPS01-USB
USB power plug Type FSP007-P01P (manufacturer’s designation)
Primary
100 – 240 V~, ±10%, 50 – 60 Hz, ±5%, ≤ 0.2 A
Secondary 5.2 V
DC
, ± 5%, 1.4 A (max.)
Further data Protection class II
IP40 in accordance with EN 60529/IEC 60529