EasyManua.ls Logo

Sartorius MIS1 - A;D Converter Verifiable Configuration Examples; Example 3: Verifiable Config, Single Range; Example 4: Verifiable Config, Multi-interval

Sartorius MIS1
107 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Example 3:
Enter or change values for verifiable configuration in single range
mode in the unit set under 1.7.x.
Select menu item ADC CON
) Confirm menu item ADC CON
(if necessary press and select the menu item VERIF.
)k p)
Verifiable configuration
) Confirm menu item VERIF.
Accuracy class
(if necessary press ) ) () To confirm accuracy class 3/4
Range selection
) Confirm menu item RANGES
(if necessary press k repeatedly) Select menu item SINGLE
Multi-interval scale
( Confirm menu item SINGLE
Weights
) Confirm menu item WEIGHTS
Verifiable scale interval
) Confirm menu item E
( ) k p Enter a value (e.g. 0.002 kg)
(if necessary press ) repeatedly)
until display E appears
k Select menu item MAX.
Enter values for the following in
the same way:
Maximum capacity (e.g. 30 kg)
Continue as shown in example 1
after entering the maximum
capacity
max.
E
Weights
SINGle
Ranges
Class
Verif.
ADC-CON
Example 4:
Enter or change values for verifiable configuration in a multi-inter-
val scale in the unit set under 1.7.x.
(the same applies for multiple range mode).
Select menu item ADC CON
) Confirm menu item ADC CON
(if necessary press
)k p) and select the menu item VERIF.
Verifiable configuration
) Confirm menu item VERIF.
Accuracy class
(if necessary press ) ) () To confirm accuracy class 3/4
Range selection
) Confirm menu item RANGES
(if necessary press k repeatedly) Select menu item MULT. INT
Multi-interval scale
( Confirm menu item MULT. INT
Weights
) Confirm menu item WEIGHTS
Verifiable scale interval
) Confirm menu item E
( ) k p Enter a value (e.g. 0.002 kg)
(if necessary press ) repeatedly)
until display E appears
k Select menu item RANGE 1
Enter values for the following in
the same way:
Range limit 1 (e.g. 6 kg)
Range limit 2 (e.g. 15 kg)
Maximum capacity (e.g. 30 kg)
Continue as shown in example 1
after entering the maximum
capacity
Range 1
E
Weights
Mult.Int
Mult.Int
Class
Verif.
ADC-CON