Sartorius QC34EDE-S - Adjustment, Calibration, and Linearization; External Adjustment and Calibration; External Linearization Procedure

Sartorius QC34EDE-S
148 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Press =/z and hold
for 2 seconds:
Place required weight on the
pan: 5 kg/10 kg
(for the required weight,
see Specifications
in the manual).
External adjustment/
calibration (code 1 9 1)
External linearization
(code 1 9 5)
Unload weighing pan.
Tare the scale.
Press =/z: 0.0 g
or 0 kg.
Adjustment/Calibration/Linearization