2. Adjustment using the PSION server
Switch the scale to the BPI mode (see page 11).
Switch on the PSION server and call the CAS program (version 4.4 or
newer)
Select the menu item “Electr-zero-point” from the “Diagnosis” submenu
With no load on the weighing pan, a value of 14.8 % ±11% should
appear in the PSION display.
If the value displayed is outside the tolerance limits, you can adjust this
voltage by changing the solder bridges LB7 and LB8.
LB7 LB7/LB8 LB8 LB7/LB8
Zerp Point Offset Value 1 / 2 1 / 1 1 / 2 2 / 2
< 25,8 % open close open open
> 3,8 % close open close open
After adjusting the zero point offset, proceed with the adjustment of the span
(sesitivity) (with the PSION server, if desired).
Adjusting the Linearity
Note: There are two methods for adjusting the linearity:
1. As described in this manual, or
2. Using the MC1 PSION server
Test the linearity of the scale in 1000g steps over the entire weighing
range.
If any deviation is greater than the tolerance values given, you need to
adjust the linearity.
Set the code “1 9 5” in the scale operating menu (external
linearization accessible).
When 0.0g is displayed, press the e key and hold it for about 3
seconds; then release it.
The first linearization value is displayed.
Place the prompted weight on the middle of the weighing pan.
After the weight is stored, the next weight value is displayed.
Each linearity support point is called and stored.
When this routine ends, the scale returns to the normal operating mode.
If desired, reset the menu item “1 9 “ to the standard setting.
Note: After linearization, make sure you perform the adjustment of the span
(sensitivity).