21
SARTORIUS QC - Scales
Lay the scale on its side.
Remove the cover plate (209) from the holder (210) (A).
Remove the strain gauge system connecting wire from the bottom
plate (106) (remove the strain relief (105)).
Unscrew the strain gauge system (103) from the bottom plate (106).
Unwind the strain gauge system connecting wire from the bottom plate
and remove the strain gauge system.
Place the new strain gauge system (103/113) on the bottom plate (106)
and screw it down loosely.
Position the strain gauge system (103/113) and tighten the fastening
screws to 25 Nm.
Attach the strain gauge system connecting wire to the bottom plate (106)
(re-attach the strain relief).
Loosely attach the screws that hold the pan support (102) on the strain
gauge system, then position it and tighten the screws (25 Nm).
Now pull the strain gauge system connecting wire into the display unit
holder.
Thread the strain gauge system connecting wire through the opening in
the rear panel of the display unit (104) and put the washer (104.C) in
place.
Fasten the strain gauge system connecting wire (and the data output
cable,
if any) with the cable clamp (103).
Attach the rear panel of the display unit (104) to the display unit
holder (108) with the two retainers (104.B).
Replace the front panel with the main PCB.
Now perform a through check and, where necessary, readjustment of the
following points:
Overload stops
Zero point offset value
Linearity
Span (Sensitivity)
The scale is now ready for operation.