User’s Manual ScanWave ● J05451 ● V5 ● (13) ● 05/2013 ● NG39EN020E ● 5
screen displays the message «Lb» (Low battery), the
handpiece beeps 4 times, and the start of a curing
cycle will be inhibited.
• You must then replace the ScanWave on its charger.
1.6 recharging
The translucent «Neck» of the charger (fig. 1-10) shows
the charger operation status and the level of lighting
power emitted by the light.
Place the charger on a hard, stable surface with a
flatness of less than 5°, then plug the connector into
the grid supply. The translucent part of the support
successively emits three flashes: red, green and blue.
Then insert the light guide in the charger and ensure
charging terminals are in contact with battery
connections.
The translucent part then emits a blue light signalling
that the battery is charging. The blue light goes out
when charging is complete.
1.7 checking lighting Power
The charger can be used to check the lighting power
emitted by ScanWave. To do this, a control window is
located in the rear section of the support (fig. 1-11).
To check the lighting power emitted:
• Select Scan mode; continuous; 9 seconds.
• Clean the optical guide, make sure that it is not
damaged, and then insert it into the handpiece.
• Also make sure that the control window is unbroken
and clean.
• Place the tip of the optical guide flat on the control
window and activate ScanWave.
• If the transluscent «Neck» of the charger lights up
green, the lighting power is sufficient. If the lighting
power is not sufficient, this will be signalled by a
fixed red or flashing red/green light. In this case,
refer to chapter X.
1.8 technical sPecifications
Model name ScanWave
Manufacturer SATELEC
Medical category IIa according to the
directive 93/42/CEE
Handpiece
Weight 185g
Dimensions Ø34 x 220mm
Operating cycle
See II - Routine operation
-
Clinical care
Category Type B
Protection 1,5 A T Fu1 fuse (non
accessible) 125V
Protection index IPX0
AC power adapter
Input voltage 100V AC to 240V AC
Frequency 50Hz to 60Hz
Output voltage 12 V DC
Output current 0.8 A
Category II
Protection index IP 41
Charging base
Input voltage 12V DC
Protection 3 A T Fu1 fuse (non
accessible) 125V
Operating cycle Continuous operation
Protection index IPX0
Battery
Type Lithium-Ion
Dimensions Ø34 x 115mm
Capacity 2500mAh
Optical specifications
LED sources 4
Wavelength range 390nm-505nm
Central wavelength 405nm / 440nm / 460nm
and 480nm
Intensity
1500 mW/cm² ± 10% (for a 7.5mm fiber : active
diameter 6.8)
2200 mW/cm² ± 10% (for a 5.5mm fiber : active
diameter 4.8)
Maximum exposure time
60 seconds (pulse mode at 1500 mW/cm² ± 10%)
Laser aiming system
Typical wavelength 655nm