Schick CDR PanX
Error 18 troubleshooting guide
Check normal operation. If error 18 appears then continue
Power Supply
Board
1) Switch
machine OFF.
2) Remove top
cover
3) Check fuses
FU304 and
FU305 on the
Main power
supply board
and replace as
necessary.
4) If fuses were
replaced then
turn machine
ON and retest
system. If the
error is not
present then
take a test
image to check
for X-ray
emission and
whether sensor
recalibration is
necessary.
5) If error still
appears or
FU304 and
FU305 are good
then check the
voltage at
connector
XC303.
Loosen (do not remove) this screw
to unclamp to top cover.
Slide the cover toward the front
and lift to remove
Check and replace fuses
FU304 and FU305 as
necessary
Caution: Dangerous Voltages are present
during parts of this testing. Please observe all
applicable precautions for High Voltage
Systems. In addition parts of this test include
emission of ionizing radiation. Please use
shielding.