EasyManua.ls Logo

Seg HighTECH Line - Page 28

Seg HighTECH Line
42 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
28 TD_MRI1_06.05_GB
6.4.7 Example of a test circuit for MRI1
relay with directional feature
Figure 6.2: Test circuit
For testing relays with directional feature, current and
voltage input signals with adjustable phase shifting are
required. Figure 6.2 shows an example of a single
phase test circuit with adjustable voltage and current
energizing the MRI1 relay under test.
For testing a relay with directional feature, one of the
input energizing quantity (voltage) shall be applied to
the relay with a constant value within its effective
range. The other input energizing quantity (current)
and phase angle shall be appropriately varied.
MRI1 is a three phase directional time overcurrent re-
lay with relay connection angle of 0°. The relay input
currents and their corresponding reference voltages are
shown in the following table (refer to 4.3):
If the single phase test circuit as illustrated in
Figure 6.2 is applied to test the directional feature of
the relay and the current source is connected to
phase 1 current input (B3/B4), then the voltage source
should be connected to relay terminals A5/A7.
In order to test the directional feature, all activation
points should first be set to “EXIT”. Then a test voltage
equivalent to the rated voltage is connected to termi-
nals A5/A7 and a current of 1 x In is impressed upon
the current inputs B3/B4.
It is now possible to read and check all measured val-
ues in accordance with Table 6.1. If the phase posi-
tion is changed, the values I
Q
and I
P
change. If the an-
gle is changed by 90
o
, for example, the measured va-
lue for current input I1 must be 1.0 for I
P
and +/- 0.0
for I
Q
.
Determining the change in direction
The angle of greatest sensitivity for determining the
phase direction is adjustable between 15
o
and 83
o
.
Consequently, the greatest sensitivity is achieved with
setting 49
o
if the input current leads the input voltage
by 49
o
. This setting results in a tripping range in ad-
vance direction of 139
o
leading to 41
o
lagging if the
marginal regions are neglected on account of lack in
measuring precision.