Storage and Transportation Temperature Limit: -20~55℃"(-4131)
SPECIFICATIONS
Type:
Digital thermometer (not predictive)
Measurement range:
32.042.9℃"(90.0109.9) ( / choosen by
manufacturer)
Precision:
±0.1(±0.2) during 35.5~42.0(95.9~107.6) at
18~28 (64.4~82.4 ) ambient operating range
±0.2(±0.4) for other measuring and ambient operating
range
Operating mode :
Direct mode
Display:
LCD, 3 ½ digits
Memory:
For storing the last measured value
Battery:
1x 1.5 V DC. button battery (size LR41or SR41, UCC 392)
Battery life:
Approx. 200hours of continuous operation or 1 year with 3
measurements per day
Dimension:
13.9cm x 2.2cm x 1.2cm (L x l x h)
Weight:
Approx. 12 grams including battery
Expected service life :
3 years
Ambient operating range:
Température : 540℃"(41104)
Humidité : 15%95%RH
Pression atmosphérique : 700hPa1060hPa
Storage and transportation
condition:
Temperature: -2055 (-4131)
Humidity: 15%95%RH
Atmospheric Pressure: 700hPa1060hPa
Ingress Protection Rating:
IP27
Classification:
Type BF
/ SWITCHABLE
Temperature readings are available in the Celsius or Fahrenheit scale (/; located in the
upper right corner of LCD.) With the unit off, press and hold the On/Off Button for
approximately 2 seconds to change the current setting.
DIRECTIONS
1. Press the On/Off Button next to LCD display. A tone will sound as the screen shows
, followed by last recorded temperature. After showing the self-test
temperature, the thermometer is now in the testing mode.
2. Position thermometer in desired location (mouth, rectum, or armpit.)
a. Oral Use: Place thermometer under tongue as indicated by
“√” position shown in Figure 2. Close your mouth and
breathe evenly through the nose to prevent the
measurement from being influenced by inhaled/exhaled air.
Normal temperature between 35.7 and 37.3℃"(96.3 and
99.1).
-20°C
55°C