A) Sensor application using MAR/e-MI or MAR/e-SF
https://youtu.be/
g3aQbOqge5Y
Measurement
Sites
1
Ensure SDM
displays “Ready
for use”.
I
s
o
p
r
o
p
a
n
o
l
7
0
%
2
Clean
Measurement
Site selected by
the home care
provider, using
Isopropanol 70%
soaked wipe.
3
Pull adhesive tape
off MAR/e and
attach MAR/e to
measurement site,
pressing gently
around the ring.
4
Open Docking
Station, remove
sensor, close
Docking Station.
Note: Do not
touch the sensor
surface with any
hard objects
or fingernails.
Otherwise, the
sensor could be
damaged.
5
Apply one drop
of Contact Gel to
the sensor surface
Note: Avoid
contact with
injured skin. Do
not use when
exhibiting allergic
reactions to one of
the components.
6
Click sensor into
MAR/e and rotate
sensor to spread
Contact Gel.
7
Secure cable with
an adhesive tape.
Important
Note: Route the
sensor cable
properly (e.g.
with a Clothing
Clip) to avoid
entanglement or
strangulation.