SPS-1500 Troubleshooting
14
Error
Possible Cause
Troubleshooting Step
No Signal
(SPS measures gauge
current but no modulation
current)
Possible SPS/gauge firmware
corruption
Possible SPS/gauge component
failure
No-Tool Current setting incorrect
in SPS
Use DC Ammeter to measure DC current from SPS to gauge.
Gauge current is normally modulated from approximately 15mA
to 25mA. If not modulated, check SPS and connections.
No Signal
(SPS mide la corriente del
sensor pero no la corriente
de modulación)
Posible corrupción del firmware
del sensor / SPS
Posible falla en algún
componente del sensor / SPS
Configuración de No-Tool
Current incorrecta en SPS
Utilice un amperímetro de DC para medir la corriente DC de SPS
al sensor. La corriente del sensor normalmente se modula
desde aproximadamente 15 mA a 25 mA. Si no está modulado,
verifique SPS y conexiones.