Cleaning & Maintenance
DO NOT USE CHAIR UNTIL YOU HAVE FULLY READ THE INSTRUCTIONS
CLEANING & DISINFECTION of ELECTRICAL COMPONENTS
The products can be cleaned as follows according to their IP protection, which is stated on the product label:
IP PROTECTION CLEANING INSTRUCTIONS
IPX0 Clean with a damp cloth.
IPX1 Clean with a damp cloth.
IPX2 Clean with a damp cloth.
IPX3 Clean with a damp cloth.
IPX4 Clean with a damp cloth.
IPX5 Wash with a brush and water, but not water under pressure.
IPX6 Wash with a brush and water. The water can be under pressure, but the system must not
be hosed down directly with a high pressure cleaner. Max. 20°C.
IPX6 W* Clean by the use of water tunnels
X can have a value of between 1 and 6.
* The following products can be marked with IPX6, but they are still allowed for wash tunnels: ACP, ACM (mini), ACM
(new), BA18, CB12/14/18, DJB, FS2, HB40, LA28, LA32, LA34 stand, LA34 comp.
The materials are resistant to the majority of cleaners and disinfectants used in the hospital and nursing home
sector. Cleaning with a steam cleaner, for example, is not permitted, as it will not be possible to maintain a minimum
distance of 300 mm from the electrical parts.
IP 66W products can be washed in wash tunnels that wash according to the “Machine Decontamination” provided
that the following guidelines are complied with:
» The cleaning and disinfection cycle in the wash tunnel must not last longer than 10 minutes.
» The nozzle pressure in the wash tunnel must not exceed 10 bar.
» The distance between the nozzle and the electrical parts must be at least 300 mm.
» The handset and motor plug must be pushed right in.
» The water temperature must not exceed 85°C.
» Cooling with cold water is not permitted.
To avoid degreasing of the piston rod, the actuator should be retracted to minimum stroke before washing.