    What to do if the Shimadzu Control Unit valve is leaking?
    • P
      pmoraAug 12, 2025
      If the Shimadzu Control Unit valve is leaking, first check if any flow line connectors are loose and tighten the male nut. If the leak persists, replace the male nut and ferrule. Also, ensure that the pressure does not exceed 34.3MPa. If the issue continues, the rotor or stator may be worn and require replacement.
  • K
    Kimberly HarrisAug 19, 2025
    How to fix a Shimadzu Control Unit that does not turn on?
    • R
      robert69Aug 19, 2025
      If the Shimadzu Control Unit does not turn on, even after switching it on, check the following: * Ensure the power plug is correctly connected. * If the power cord's internal wires are cut, replace it with a new cord of the same type. * Verify that the power supply meets the instrument's specifications. * Check if the fuse is blown and replace it if necessary.
  • G
    Gina WilliamsAug 27, 2025
    Why is the Shimadzu Control Unit valve inlet pressure high?
    • S
      Sherry RuizAug 27, 2025
      If the Shimadzu Control Unit valve inlet pressure is high, the valve flow lines may be clogged. Try performing a flow line cleaning. If the issue persists, the valve switching position may be misaligned, and you should contact your Shimadzu representative.
  • M
    Mary BerrySep 3, 2025
    What to do if Shimadzu Control Unit valve does not move when performing contact input?
    • D
      Diamond KimSep 3, 2025
      If the Shimadzu Control Unit valve does not move when performing contact input, it could be due to several reasons. * First, ensure that remote mode is enabled by pressing [local/remote]. * Next, verify that the event cable is correctly connected. * Finally, check if the event cable is cut and replace it if needed.