4. Shut off the exhaust ventilation system and shut off argon gas at the source.
5. Shut off helium for the collision cell at the source.
6. Ensure that the chiller water has antimicrobial additives as recommended in
the user manual. Three tablets of baby aspirin dissolved and circulated can
act as a substitute for commercially available products. Shut off Chiller.
7. Shut down the LabSolutions ICPMS software and the control computer.
8. Shut off power to the ICPMS, autosampler, and other external accessories.
9. Turn off external roughing pump.
10. Dispose of the rinse solution in the J-tube and drain line and close the waste
collection container.