ICPMS Preparing your instrument for shutdown
1. To flush out the system, aspirate with rinse solution for five minutes then DI
water for an additional three minutes.
2. Shut down the plasma by clicking the Plasma Off button in LabSolutions
ICPMS software at the bottom-left of the screen. Wait for the shutdown
sequence to complete.
3. Release tension on the peristaltic pump tubing by releasing the guide arms
and detaching tubing from the tube hangers.
Page 1 of 4
Question and Answer IconNeed help?

Do you have a question about the Shimadzu ICPMS and is the answer not in the manual?

Summary

ICPMS Shutdown Procedure

System Flushing and Plasma Off

Flush the system with rinse solution and DI water, then shut down plasma via software.

Peristaltic Pump Tubing Release

Release tension on peristaltic pump tubing by detaching it from tube hangers.

Gas and Chiller Shutdown

Shut off exhaust ventilation, argon, and helium. Ensure chiller additives and turn off chiller.

Software and Computer Shutdown

Shut down LabSolutions ICPMS software and the control computer.

Power and Pump Shutdown

Turn off power to ICPMS, autosampler, and accessories. Turn off external roughing pump.

Waste Disposal

Dispose of rinse solution from J-tube and drain line, then close waste container.

Extended Shutdown Instructions

For shutdowns >2 months, follow user manual instructions regarding backup battery.

Stop of the Equipment

Check backup battery level before stopping. Replace if instrument is off for >2 months.

ICPMS Startup Procedure

Refer to Instruction Manual

Consult the ICPMS-2030 Instruction Manual for specific startup procedures.

Duct Exhaust and Cooling Water

Start duct exhaust system and supply cooling water after checking levels.

Roughing Pump and Gas Supply

Check roughing pump oil, turn on power, then turn on argon and helium sources.

Instrument and PC Power On

Turn on the main ICPMS power switch, then the PC and peripherals.

Vacuum Pump Down

Allow vacuum to pump down overnight before turning on the plasma.

Pressure Seal and Checks

Fill J-tube with water for pressure seal. Check glassware, cones, and pump.

Summary

ICPMS Shutdown Procedure

System Flushing and Plasma Off

Flush the system with rinse solution and DI water, then shut down plasma via software.

Peristaltic Pump Tubing Release

Release tension on peristaltic pump tubing by detaching it from tube hangers.

Gas and Chiller Shutdown

Shut off exhaust ventilation, argon, and helium. Ensure chiller additives and turn off chiller.

Software and Computer Shutdown

Shut down LabSolutions ICPMS software and the control computer.

Power and Pump Shutdown

Turn off power to ICPMS, autosampler, and accessories. Turn off external roughing pump.

Waste Disposal

Dispose of rinse solution from J-tube and drain line, then close waste container.

Extended Shutdown Instructions

For shutdowns >2 months, follow user manual instructions regarding backup battery.

Stop of the Equipment

Check backup battery level before stopping. Replace if instrument is off for >2 months.

ICPMS Startup Procedure

Refer to Instruction Manual

Consult the ICPMS-2030 Instruction Manual for specific startup procedures.

Duct Exhaust and Cooling Water

Start duct exhaust system and supply cooling water after checking levels.

Roughing Pump and Gas Supply

Check roughing pump oil, turn on power, then turn on argon and helium sources.

Instrument and PC Power On

Turn on the main ICPMS power switch, then the PC and peripherals.

Vacuum Pump Down

Allow vacuum to pump down overnight before turning on the plasma.

Pressure Seal and Checks

Fill J-tube with water for pressure seal. Check glassware, cones, and pump.

Shimadzu ICPMS Specifications

General IconGeneral
TypeInductively Coupled Plasma Mass Spectrometer (ICP-MS)
ApplicationsEnvironmental monitoring, pharmaceutical analysis
Detection LimitsSub-ppt levels (parts per trillion) for many elements, depending on the matrix and element.
Plasma SourceArgon-based inductively coupled plasma
Sample Introduction SystemNebulizer, spray chamber
DetectorElectron Multiplier
ModelMultiple models available (e.g., ICPMS-2030, ICPMS-2050)
RF Generator27.12 MHz free-running solid-state generator
InterfaceSampler and skimmer cones
Vacuum SystemTurbo molecular pump backed by rotary pump
Data AcquisitionFast scanning
Power Requirements200-240 V, 50/60 Hz
WeightApproximately 150-200 kg (varies by model)