Flush the system with rinse solution and DI water, then shut down plasma via software.
Release tension on peristaltic pump tubing by detaching it from tube hangers.
Shut off exhaust ventilation, argon, and helium. Ensure chiller additives and turn off chiller.
Shut down LabSolutions ICPMS software and the control computer.
Turn off power to ICPMS, autosampler, and accessories. Turn off external roughing pump.
Dispose of rinse solution from J-tube and drain line, then close waste container.
For shutdowns >2 months, follow user manual instructions regarding backup battery.
Check backup battery level before stopping. Replace if instrument is off for >2 months.
Consult the ICPMS-2030 Instruction Manual for specific startup procedures.
Start duct exhaust system and supply cooling water after checking levels.
Check roughing pump oil, turn on power, then turn on argon and helium sources.
Turn on the main ICPMS power switch, then the PC and peripherals.
Allow vacuum to pump down overnight before turning on the plasma.
Fill J-tube with water for pressure seal. Check glassware, cones, and pump.
Flush the system with rinse solution and DI water, then shut down plasma via software.
Release tension on peristaltic pump tubing by detaching it from tube hangers.
Shut off exhaust ventilation, argon, and helium. Ensure chiller additives and turn off chiller.
Shut down LabSolutions ICPMS software and the control computer.
Turn off power to ICPMS, autosampler, and accessories. Turn off external roughing pump.
Dispose of rinse solution from J-tube and drain line, then close waste container.
For shutdowns >2 months, follow user manual instructions regarding backup battery.
Check backup battery level before stopping. Replace if instrument is off for >2 months.
Consult the ICPMS-2030 Instruction Manual for specific startup procedures.
Start duct exhaust system and supply cooling water after checking levels.
Check roughing pump oil, turn on power, then turn on argon and helium sources.
Turn on the main ICPMS power switch, then the PC and peripherals.
Allow vacuum to pump down overnight before turning on the plasma.
Fill J-tube with water for pressure seal. Check glassware, cones, and pump.
| Type | Inductively Coupled Plasma Mass Spectrometer (ICP-MS) |
|---|---|
| Applications | Environmental monitoring, pharmaceutical analysis |
| Detection Limits | Sub-ppt levels (parts per trillion) for many elements, depending on the matrix and element. |
| Plasma Source | Argon-based inductively coupled plasma |
| Sample Introduction System | Nebulizer, spray chamber |
| Detector | Electron Multiplier |
| Model | Multiple models available (e.g., ICPMS-2030, ICPMS-2050) |
| RF Generator | 27.12 MHz free-running solid-state generator |
| Interface | Sampler and skimmer cones |
| Vacuum System | Turbo molecular pump backed by rotary pump |
| Data Acquisition | Fast scanning |
| Power Requirements | 200-240 V, 50/60 Hz |
| Weight | Approximately 150-200 kg (varies by model) |