Reuse
This medical device can be reused by another owner if the original owner of the product doesn’t
need it anymore. Before reuse the device must be carefully inspected by authorized technical
staff.
Following elements should be checked:
• rigidity of the structure,
• tightness of bolts and nuts,
• condition of plastic elements,
• condition of wheels and ball bearings,
• efficiency and condition of brakes.
If
any of above elements are torn or broken, they should be replaced with new parts. It is specially
recommended to install new wheels, seat and backrest for every new user of the device. All lose
bolts and nuts must be tightened. The product must be precisely cleaned and disinfected. Every
new owner must receive the device with this user’s manual attached.
WARNING! It is forbidden to reuse the d
evice if the main frame is bent or broken.
Disposal of Your device
The device should not be deposited with normal garbage. Your shower chair must be disposed
according to applicable local and national statutory regulations. Contact your local waste disposal
agency or authorized dealer of rehab equipment for information on proper disposal of the chair,
packaging, metal frame components, plastic components a
nd other materials.