-13-
Line up the small white bulge on the
Sensor Applicator with the concavity on
the edge of Sensor Pack. On a hard
surface, press firmly down on the Sensor
Applicator until it comes to a stop.
CAUTION: DO NOT remove the safety
clip (see step 6) from the Sensor
Applicator prior to pressing into the
Sensor Pack. If the safety clip is removed
prior to this assembly, damage may be
caused by accidentally pressing the
button that inserts the Sensor
prematurely.
Lift the Sensor Applicator out of the
Sensor Pack.
CAUTION: The Sensor Applicator now
contains a tip. DO NOT touch the inside
of the Sensor Applicator or put it back
into the Sensor Pack.
Press the safety clip and pull it out from
the Sensor Applicator.
Place the Sensor Applicator over the
prepared site and push down firmly to
apply the Sensor to your body.
Precaution: DO NOT push down on the
Sensor Applicator until placed over the
prepared site to prevent unintended