-36-
IMMUNITY test
IEC/EN 60601 test
level
Compliance Level
Electromagnetic
environment
guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
±8 kV contact
±2 kV, ±4 kV; ±8 kV,
±15 kV air
±8 kV contact
±2 kV, ±4 kV; ±8
kV, ±15 kV air
Floors should be
wood, concrete or
ceramic tile. If floors
are covered with
synthetic material, the
relative humidity
should be of at least
30 %.
Electrical fast
transient
IEC 61000-4-4
±2 kV for power
supply lines
±1 kV for input/
output lines
Not applicable
Not applicable
Surge
IEC 61000-4-5
±0.5 kV, ±1 kV line
to line
±0.5 kV, ±1 kV, ±2
kV line to ground
Not applicable
Not applicable
Voltage dips and
interruptions
IEC 61000-4-11
0%, 70%, 0% of U
T
Not applicable
Not applicable
Power frequency
(50/60 Hz) magnetic
field
IEC 61000-4-8
30 A/m
30 A/m
Power frequency
magnetic fields should
be at levels
characteristic of a
typical location in a
typical domestic,
commercial or hospital
environment.
NOTE: U
T
is the AC Mains voltage prior to application of the test level.