-37-
IMMUNITY test
IEC/EN 60601
test level
Compliance level
Electromagnetic
environment – guidance
Conducted RF
IEC 61000-4-6
150 kHz to 80
MHz 3V
ISM and amateur
radio bands
between 150 kHz
to 80 MHz 6V
Not applicable
Not applicable
Radiated RF
IEC 61000-4-3
80MHz to
2700MHz
10V/m
385MHz 27V/m
450MHz 28V/m
710MHz,
745MHz,
780MHz 9V/m
810MHz,
870MHz,
910MHz 28V/m
1720MHz,
1845MHz,
1970MHz
28V/m
2450MHz
28V/m
5240MHz,
5500MHz,
5785MHz 9V/m
10V/m, 80% Am at
1kHz
27V/m PM at 18Hz
28V/m FM Β± 5 kHz
deviation at 1kHz
sine
9V/m PM at 217Hz
28V/m PM at 18Hz
28V/m PM at
217Hz
28V/m PM at
217Hz
9V/m PM at 217Hz
d = 1.2
√
𝑃
80 MHz to 800 MHz
d = 2.3
√
𝑃
80 MHz to 2.5 GHz
P is the maximum output power rating of the transmitter in watts (W) according to the
transmitter manufacturer and d is the recommended separation distance in meters (m).
Field strengths from fixed RF transmitters, as determined by an electromagnetic site survey,
should be less than the compliance level in each frequency range.
Interference may occur in the vicinity of equipment marked with the following symbol:
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is
affected by absorption and reflection from structures, objects and people.