Only Trained Healthcare Professionals are
Authorized to use the Sigma Spectrum Pump
sigma spectrum
Quick Reference Guide
2
3
4
2
3
4
Starting a New Infusion
Press POWER ON KEY wait for power-up to complete.
1.Gravity prime the IV, removing all air from Y sites
and backcheck valve. Position and close the roller
clamp on the IV set comfortably below the pump.
2.Insert the slide clamp into the keyhole on top of the
pump at .
3.Push down on the slide clamp until the door opens.
4.Load the tubing into the tubing guides as illustrated.
Starting at the top with the slide clamp at
proceeding in order through guides &
ensuring the fluid flow is directed from the IV
container to the patient.
5.Close the door by pressing with both thumbs at top
and bottom of door over latches.
6.The IV set is properly loaded when the screen
displays three green bars and check marks as
illustrated in Fig. 1, on page 2.
Set Loading
2 3 4
ON
OFF
14
MFGID
sigma, LLC
711 Park Avenue
Medina, NY 14103
info@sigmapumps.com
www.sigmapumps.com
Troubleshooting/Operating Tips
Soft keys
ŸAlarms - Follow all prompts on screen to clear alarms. Silence will be the
right soft key. Press SILENCE soft key or any key to silence the alarm for 2
minutes.
ŸStart with a fresh section of tubing loaded in the pump, make sure there is
enough tubing for patient movement.
ŸCorrect loading is confirmed by drops falling in drip chamber after pressing
RUN. Confirm flow from IV container, to drip chamber, to patient.
ŸAlways press OK to confirm entries and changes.
ŸKeep the power cord plugged in whenever the pump is not in use.
ŸOptions Allows operator adjustment of alarm and display settings.
ŸTitrate Press the TITRATE soft key to change the flow rate and VTBI
(volume to be infused) without stopping the pump.
ŸReview Displays infusion information.
ŸReview/Program Return to primary setup screen.
ŸHold Infusion is in stand-by (No alarm ) until any key is pressed.
P/N 95520-605 REV A
From IV
Container
From IV
Container
To PatientTo Patient
Call 1-800-356-3454
24-7 Technical Support
Call 1-800-356-3454
24-7 Technical Support

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Sigma spectrum and is the answer not in the manual?

Sigma spectrum Specifications

General IconGeneral
TypeInfusion Pump
DisplayColor LCD
ModelSpectrum
Infusion Rate0.1 to 1200 mL/hr
Power SupplyAC/DC
AlarmsAir-in-line, occlusion, low battery, end of program
CompatibilityBaxter Administration Sets