INST‑05070410
Section 8. TABLES OF FUNCTIONS
8.1 Function switches
Situation Function Color Operations Mode Unit A d di ti on a l
switch display
Pressing less than 3 seconds Standard
kV
POWER A.OFFON/OFF RED Pressing 3 seconds or more Continuous
Press once more to turn off −−
IB V IBBLUE Ion balance measurement Standard or
ON/OFF. continuous
After
HOLD HOLDPower GREEN Hold function ON/OFF. Holds the digital value and the
ON bar graph information by pressing once.
Zero adjustment of the displayed values by pressing once. The display
GRAY should read less than [±0.30] kV for static charge measurement and lessZERO
than [±50] V for ion balance measurement for zero function to operate.
8.2 Beep sounds
Once Three times Twice for five times each at Continuous
one second interval
「」 Pi PiPiPi PiPi Piii
0.25 seconds ON 0.15, OFF 0.1 seconds ON 0.15, OFF 0.1 seconds
When power is turned on. When power is turned on. 5 seconds before stopping M e a s u r i n g
POWER button is pressed POWE R button is pressed for by auto power off function. range exceeded
for less than 3 seconds. 3 seconds or more.
8.3 Red LEDs for focussing
Static charge measurement Ion balance
measurement
By pressing POWER button for When HOLD During is shownIB
button is pressed
Less than three seconds Three seconds or more once.
until power It turns off after one minute duringLight ON Light ON. Light OFF
Light OFFis turned off. automatically. HOL D or ZERO button the time, isHOLD
can be used to turn on again. displayed.
8.4 White LED to assist in reading display
Pressing time o f POWER button T 1 s 1 s T 3 s 3 s T 5 s T 5 s<≦
when power is turned on (T)
OFF ON OFF ONLED situation
-page20-
INST‑05070410
Section 9. TROUBLESHOOTING
Trouble Cause Countermeasure
The digital display and the bar
graph on LCD do not change. The meter is in hold Press HOLD button once again for canceling
mode. hold mode. Confirm that "HOLD" is not
Even if ZERO button is showed on the display.
pressed, the display cannot be
set to ze ro. The true zero point is If the meter reads 0.30 kV or 50 V or more
shifted. when power is turned on toward an uncharged
object, is needed.readjustment
LCD shows nothing or a part No battery or low Put in a battery or replace the battery as
is missing. battery required.
Malfunction of the LCD is needed.Repair or replacement
White LED to assist in reading The pressing time of Refer to section 8.4.
display does not work. POWER button is not
correct.
FMX-003 gives a reading FM X-00 3 is turned Turn toward an uncharged object.
different from zero during zero toward a charged object.
checking.
Bad grounding or the Ground FMX-003 using the grounding lead.
operator is charged.
Display shows "Err". Malfunctio n o f the of the sensor isRepair or replacement
sensor. needed.
Beep sound does not activate. Malfunction of the of the buzzer or PCBRepair or replacement
buzzer or the circuit. is needed.
Indicated voltage is high in ion It is not in ion balance Press IB button once. Confirm that "IB" is
balance mode. mode. shown on the display and the unit changes to
"V".
Indicated voltage is low in ion Insulation resistance of Clean the plate by u ltrasonic an d dry
balance mode. the Ion balance plate is completely in a desiccator or in a poly bag
low because of adhesion using desiccant.
of dirt or condensation
on the plate insulation. is needed, if noRepair or replacement
improvement is noticed.
If the countermeasures fail or readjustment, repair or replacement is needed (except battery), do not
use any more and please contact SIMCO JAPAN with the product's serial number and information
on the failure.
-page21-