EasyManua.ls Logo

Sitron CF12 AC - CF12 Calibration Procedures; CF12 Calibration Setup; CF12 Calibration for Flow;No Flow; CF12 Set Point Adjustment

Sitron CF12 AC
20 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
13
Calibration
To Start:
1 - Remove the enclosure lid (Note: the screws are self-retaining)
2 - Start the power supply and wait 5 minutes until the CF12 is active and has reached a
stable point within the medium.
3 - Let the regular or desired flow reach its point of normal operation.
Calibration for Flow / No Flow:
1 - Set the flow rate at the normal range of operation.
2 - Turn the potentiometer counter-clockwise until the central LED turns red.
3 - With the central LED red, turn the potentiometer clockwise until the central LED
changes to a blinking green state.
4 - Continue to turn the potentiometer clockwise until all green LEDs in the bar graph are
on.
5 - In order to be sure that the adjustment is not at a critical state give the potentiometer
an additional quarter turn clockwise.
Set Point Adjustment:
The flow switch can be adjusted to indicate either increasing flow, or decreasing flow at a
specific set point within 3cm/s to 3m/s.
It is important to determine the specific set point at which the flow switch should activate
or de-activate.
0%
100%
BARGRAPH - Lights ON
Set Point
BARGRAPH - Lights ON
0%
100%
BARGRAPH - Lights OFF
Set Point
BARGRAPH - Lights OFF
Adjust
Adjust