Troubleshooting
and Care
43
You seeDVORZO\ŴDVKLQJ green
check mark icon and the
progress bar is frozen.
You hear a repeating warning sound.
Injection paused (Step 9 on page 30). If the green top
is removed from the skin before completing the
injection, the auto-injector will pause the injection
(see Figure AC).
Do this: Press the green top against skin and the
injection will start again. If the injection is not completed
within 15 minutes, the auto-injector automatically
cancels the injection.