OPERATION
32
conjunction with the Key Lock. Within this selection, you can lockout all front panel
control functions except TEST, RESET and memory select.
You can toggle this selection only from a power OFF state. Please turn the POWER
switch to the off position then press and hold the SET key, then turn the power ON.
Momentarily you will see one of the following screens to indicate if the keys are locked
or unlocked:
or
To toggle the Key Lock function on and off use the Up or Down Arrow keys, then press
EXIT to select, or press the SET key again to reach the Memory Lock screens.
The memory lock screens appear as follows:
or
To toggle the Memory Lock function on and off use the Up or Down Arrow keys, then
press EXIT to select, or press the SET key again to reach the Key Lock screens.
Use the following chart to set the Key Lock and Memory Lock for your application.
Key Lock
Memory Lock
Accessible keys
Off
Off
All
Off
On
All
On
Off
TEST, RESET, and Up-Down Arrows for memory selection
On
On
TEST and RESET
12 To set the IR output Test Voltage on model 2945
Please press the SET key until the display shows:
I - V o l t = X.XX KV
Please use the Up/Down Arrow keys to enter the desired test voltage, then press the EXIT
key to exit to the test mode or toggle to another setting using the SET key. The maximum
voltage which may be entered is 1.00 kilovolts DC.
13. To set the High Resistance Limit on model 2945
Please press the SET key until the display shows:
Key Lock = ON
Key Lock = OFF
Memory Lock = ON
Memory Lock = OFF