INFORMATION SHEET #013
Trouble Shooting Chart – SNOTTY TOO
The below chart has been created to streamline trouble shooting procedures and enable
better understanding of the functionality of SNOTTY TOO.
PROBLEM CAUSE REMEDY
Flat Battery Replace
batteries
Wet Battery Dry for 24h.
* Note SNOTTY
TOO is not water
proof
1. Aspirator not starting
Batteries installed
incorrectly
Install batteries
properly
Aspirator has not been
cleaned properly
Clean aspirator as
per user manual
2. Low suction
at collar tap (6)
If cleaned as per above
and still not working
Return for
warranty
assessment
Base plate not attched
properly
Twist plate until
lines align on top
of aspirator
3. Low suction
(at base plate side hole)
Small o-ring not sealing Adjust or
replace small
o-ring (11)
If nozzle is REMOVED
and suction is same
strength as at base plate
side hole
The aspirator is
functional. The
small hole size
at tip of nozzle
reduces the
suction
sensation.
4. Low suction
(at collection cup tip
WITH or WITHOUT
silicone tip)
If nozzle is REMOVED
and there is NO suction
Check or replace
o-ring (3)
Question and Answer IconNeed help?

Do you have a question about the Snotty TOO and is the answer not in the manual?

Snotty TOO Specifications

General IconGeneral
BrandSnotty
ModelTOO
CategoryMedical Equipment
LanguageEnglish

Summary

SNOTTY TOO Troubleshooting Guide

Aspirator Not Starting

Troubleshooting steps for when the aspirator fails to power on, including battery checks and installation.

Low Suction Issues

Diagnosing and resolving problems with reduced suction power at various points of the aspirator.