1260 Impedance Gain-Phaze Analyzer Messages and Error Codes
12600012_Gmacd/CB 9-3
9.2 ERROR CODE SUMMARY
9.2.1 GROUP 0: COMMAND STRUCTURE
MESSAGE EXPLANATION
01. UNKNOWN COMMAND Command not included in instrument command set
02. ARG MISMATCH Command contained the wrong type, or wrong number, of
arguments.
03. OUT OF RANGE Argument value out of range.
04. FORMAT ERROR Floating point format error. Attempt made to enter a floating
point number in an incorrect format, e.g. 1.2.5E2 instead 1.25E2.
05. ILLEGAL REQUEST Illegal request for parameter value. Some control parameters
have no value, e.g. HS (sweep hold).
06. INVALID FUNCTION Function syntax error.
07. NO. OUT OF RANGE Integer out of range for store or constant in scaling function.
08. INVALID SYMBOL Attempt made to enter an invalid symbol in scaling function.
09. AMPL ILL. FOR HF Attempt made to enter an amplitude >1V at a frequency
>10MHz.
9.2.2 GROUP 1: LEARNT PROGRAM
MESSAGE EXPLANATION
11. ILLEGAL EDIT After EDIT has been selected, with an En command, the only
valid commands are:
#F Select next instruction.
#B Select previous instruction.
#I Insert command(s) that follow as program instruction(s).
#D Delete presently selected instruction.
#Q Quit edit.
12. ILLEGAL COMMAND Command cannot be learnt. Some commands, e.g. En (edit),
cannot be used as learnt program instructions.
13. NO SUCH PROGRAM Program does not exist, under number specified.
14. NESTING ERROR Invalid recursion attempted. A program can execute itself only if
the execute program instruction (EPn) is the last instruction
entered before *Q (quit).
(Continued on next page.)